Case BlogDOI Number : 10.36811/ojrmi.2022.110047Article Views : 0Article Downloads : 1

A Flat Patient’s Table to Prevent Malpositioning of the Patient: An Illustrated Case Study in Radiology

Abdulwahab Alahmari*

Radiology Specialist, Radiology Department, Al-Namas General Hospital, Ministry of Health, Al-Namas City, Saudi Arabia

*Corresponding Author: Abdulwahab Alahmari, Radiology Specialist, Radiology Department, Al-Namas General Hospital, Ministry of Health, Al-Namas City, Saudi Arabia, Tel: +966562428716; Email: afaa99@hotmail.co.uk

Article Information

Aritcle Type: Case Blog

Citation: Abdulwahab Alahmari. 2022. A Flat Patient’s Table to Prevent Malpositioning of the Patient: An Illustrated Case Study in Radiology. O J Radio Med Img. 5: 31-36.

Copyright: This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium,provided the original author and source are credited. Copyright © 2022; Abdulwahab Alahmari

Publication history:

Received date: 26 July, 2022
Accepted date: 05 September, 2022
Published date: 09 August, 2022

Abstract

There is a design issue in most of the CT scanners patient table/couch that cause the patient to be in a tilted position. There for a new design must help the patient to be centrilized. The design need to make a few changes to the patent’s couch. All manufacturing companies make this mistake. This paper will address the issue in details.

Keywords: Computed Tomography; Pateint Table; Patients Position; Radiology

Introduction

There have been many design issues in radiology machines reported before[1-3]. The patient couch is design in some CT scanner as curved couch which make the patient in a malposition. The malposition can make the organs to move to the low side. Even in a minor fashion, but it is still can cause misdiagnosis. In order to use the curved patient couch, a bubble level scale must be used to make sure the patient is 100% centerilzed which is crazy. The easiest solution is to make a flat patient couch which will solve the whole issue see (Figures. 1 thru 6).

ojrmi1047_figure1

Figure 1: The curved patient table.

ojrmi1047_figure2

Figure 2: The curved patient table which cause malposition.

ojrmi1047_figure3

Figure 3: The curvature of the patients’ couch.

ojrmi1047_figure4

Figure 4: The patient position on both the flat tables (A & B) and the curved table (C & D).

ojrmi1047_figure5

Figure 5: An abdomen CT scan shows the skin line (the green arrow) and the curved table top (the red arrow). The CT shows the patient tilted to the left side.

ojrmi1047_figure6

Figure 6: A pelvis CT scan shows the urinary bladder falling to the left side (the blue arrow), skin line (the green arrow), and the curved table top (the red arrow). The patient appears to be centralized, but a little bit tilted to the left side which made the bladder only on one side which can be misdiagnosed as a uterine cyst for non-experienced health professionals.

Design Issue

This is an old Siemens CT scanner, but they still make the same mistakes in newer scanners. As well, General Electric, Philips, Toshiba, Canon, and Hitachi. All of them have the same mistake which makes the patient positioning above the table hard to do. The laser beam centering is used, but the patient still not centerilized 100%. The laser centering will be on the accurate spot, but the patient legs or lower trunk could be shifted to one side of the table. Some patient might have scoliosis which make them twisted on the couch or any other medical condition. The idea of making a curved patient table is a huge mistake. The flat table can have a radiolucent cushion to make the patient comfortable on the flat patient table. As well, in emergancy suitaions where the patient –some times when the patients are brought by the relatives, not the EMS servies- not on a hard board, then it will make the patient lie on nonstright surface which is not good for the spine in case of truma. Even with patients on hard board it will make centerilizing the hard board a difficult task. Because the table is curved and the patient could be on any part of the board. Pushing the patient on the board or sliding the patient on the board is not an option to avoid causing harm in cause of spine furcture. So you have to centrilzed the patient, not the board which will make the patient be in tileted to one side more than the other. All of these issues are facing radiographers because the patient table was poorly designed by “non- radiology experts”.

Affected Organs

There is a list of organs that can be affected with a tilted positions on the patients’ table. Some of these organs are: the brain, the stomach when it is full, the urinary bladder when it is not empty, gallbladder, the liver, the bowel, the heart, the lungs when it has fluids, etc. Basically, any organ that has some fluid, the shift will be obvious. Other organ might appear in a wrongposition due to the lack of accurate positioning. As well, in cases of pathologies like a cyst or ascites, it can appears directing to the lowest level near the table top.

Central Line

The patient couch should have a red line in the center of the couch so the patients will know where they supposed to be centralized. Many X-ray bucky have the lines to help the patients’ to be centralized. The issue is the lines might not be in a distinct color. If the line is in red color for example, all patient will see it when you ask them to be centralized on the red line in the middle of the table see (Figure 7).

ojrmi1047_figure7

Figure 7: The centralization line (red arrows) and the Automatic Exposure Control (AEC)/Ionization Chambers in (green arrows).

Similar Issues in Other Modalities

In another paper about using a flat couch in MRI scan for radiotherapy planning, it should be more accurate positioning on a flat table can help in delivering radiotherapy treatment [4]. The use of curved tables in CT scanners, are most likely to allow the patients to centralized themselves appropriately. But from a clinical experience, most of the patients throw themselves on the table without centralizing. Even when instructions are given, they still do not centralize themselves. For religious reasons, a male radiographer is not allowed to touch a female patient to centralized her in appropriate way. Touching a female patient (hands, arms, feet, any part) will be considered a form of sexual harassment and against the code of conduct in many Muslim countries. As well, the dress of the patient will play a role in positioning the patient in a wrong way (i.e., removing the dress is not allowed). Other reasons for making the table curved is to allow patient comfort and the design is done by nonradiology experts (i.e., biomedical engineering).

Conclusion

A flat table with a centralization line can help the patient to adjust his/her-self on the couch for the CT scanner. With a flat table, the fluid level will not move to one direction due to the curvature of the couch, but the patient will remain in a “status of equilibrium” on a flat table.

References

1. Abdulwahab Alahmari. 2022. Engineering and Designs Mistakes in Radiology. OJ Radio Med Img. 5: 13-17.
2. Abdulwahab Alahmari. 2022. Cold CT scanner rooms: A simple solution for the patient comfort and for hypothermia cases. O J Radio Med Img. 5: 07-09.
3. Alahmari A. 2022. A Proposal of a Head Holder for CT Scan Tables: A New Fixation Design. Austin J Radiol. 9: 1191.
4. McJury M, O'Neill A, Lawson M, et al. 2011. Assessing the image quality of pelvic MR images acquired with a flat couch for radiotherapy treatment planning. The British Journasl of Radiology. 84: 750-755. Ref.: https://pubmed.ncbi.nlm.nih.gov/21750138/ Doi: https://doi.org/10.1259/bjr/27295679

Download Provisional PDF Here

Download PDF
Indexing
Platforms
Plagiarism Checker.png
Google_Scholar.png
Publons.jpg
Cite Factor.jpg
raft_ref1.jpg
DRJI.jpg
Scilit.jpg
Semetic scholar.jpg
Worldcat.jpg
icmje.png
orcid.jpg
openaire.jpg
zenodo_raft.jpg
raftpubs-base1.jpg
Mendeley.jpg
growkudos.jpg
researchgate.jpg
Academic Microsoft.jpg
dimensions-logo-400x801.png
sciencegate.png